Braun ThermoScan 5 is fast, gentle, easy & accurate, making it great for checking for fevers in infants & kids, & the reason Braun ear thermometers are used in more pediatrician's offices than other children's thermometers. View and Download Braun IRT 4020 instruction manual online. Welcome to ManualMachine. You have been successfully registered. We have emailed you a verification link to to complete your registration. Please check your inbox, and if you can’t find it, check your spam folder to make sure it didn't end up there.
- Braun Thermoscan Instructions 4520
- Braun Thermoscan Instructions Irt 4520
- Braun Thermoscan Ear Thermometer Instructions 4520
Braun ThermoScan IRT 4520 Professional Ear Thermometer Brand New & Retail Packaged Meets the American Society for Testing and Materials (ASTM) accuracy requirement specified in ASTM standard E1965-98 as it pertains to infrared thermometers. «I/O» button (On/memory function - IRT 4520 only) 7. «ExacTemp» light 8. Battery door 10. Protective cover (IRT 4520) 11. Protective Cap (IRT 4020) How to your Braun ThermoScan 1. For accurate measurements, provide a new clean lens filter (1 Overview of the contents on page No. 7 IRT4520/4020NAS4-42.
10
Trouble-shooting
Error message
Situation
Solution
No lens filter is attached.
Attach new, clean lens filter.
The thermometer probe was not
positioned securely in the ear.
An accurate measurement was
not possible.
Braun Thermoscan Instructions 4520
POS = position error
IRT 4020: Clear the display by
pushing the Start button once.
IRT 4520: Clear the display by
pushing the «I/O» button once.
Take care that the positioning of
the probe is correct and remains
stable.
Ambient temperature is not within
the allowed operating range
(50 – 104 °F or 10 – 40 °C).
Allow the thermometer to remain
for 30 minutes in a room where
the temperature is between
50 and 104 °F or 10 and 40 °C .
Temperature taken is not within
typical human temperature range
(93.2 – 108 °F or 34 – 42.2 °C).
HI = too high
LO = too low
Make sure the probe tip and lens
are clean and a new, clean lens
filter is attached. Make sure the
thermometer is properly inserted.
Then, take a new temperature.
System error – self-check display
flashes continuously and will not
be followed by the ready beep
and the ready symbol.
If error persists,
If error still persists,
Wait 1 minute until the thermo-
meter turns off automatically, then
turn on again.
... reset the thermometer by re-
moving the batteries and putting
them back in.
… call Braun ThermoScan
customer service at
1-800-327-7226